What is an isolation transformer? How do you scope the mains side?

An SMPS repairer's most expensive mistake: clipping the scope ground onto the bridge's negative. A spark, a vaporised trace and often a dead scope input. The reason is simple: the clip is earth, and neutral is earth too.

What does it do? In the grid, neutral is bonded to earth in the panel; a bench scope's probe clip is earthed through its plug too. The 'hot GND' on an SMPS primary (the bridge's negative) connects to L through a diode during one half of the mains cycle. Clipping onto it shorts L straight to earth. A 1:1 isolation transformer feeds the device from a winding that has no link to earth: even if the clip ties one point to earth, there's no path for current to return. A differential probe measures the difference between two points without connecting to earth at all. Remember: an isolation transformer doesn't protect you from touching two points at once, and it makes any RCD behind it useless.